Electronic and Physical Specification
• Display : 7″ TFT LCD touch screen Top LED light
• Dimension : 1350mm (H) x 560mm (W) x 620mm (L)
• Weight : 85kg (Gross Weight)
• Power Input : 100 – 240V, 50/60Hz
• Power Consumption : ≤50VA
• Backup Battery : Minimum 120 min
Pneumatic Specification
• Gas Supply : O2, N2O and Reserved O2
• Yoke : 1 x O2, 1 x N2O (Optional)
• Flowmeter : O2 : 0 – 1L/min, 1-10L/min | N2O : 0 – 1L/min, 1-10L/min
• Pressure Gauge : 2 for pipelines gas, 2 for cylinder gas (Optional)
• O2 Flush : 25 – 75L/min
• Operational Modes : Closed and Semi-closed circuit system semi – open
• Driven Mode : Pneumatically driven and electronically controlled
• AGSS (Passive) : Optional
Ventilator Specification
• Patient Type : Pediatric and Adult
• Ventilation Mode : A/C, SIMV, Manual, Standby
• Tidal Volume : 20 – 1500ml
• Frequency : 0 – 100bpm
• I:E Ratio : 10:1 – 1:10
• Pmax : 20 – 70cmH2O
• PEEP : OFF, 4 – 30cmH2O (Optional)
• Pressure Trigger : 0 – 20cmH2O
• Flow Trigger : 0 – 20L/min
• CO2 Absorption System : 1500 ML
• APL Valve : 0.19 – 6KPa
Monitoring and Alarm
Monitoring
• Volumes : MV, VTe, VTi
• Pressure : Ppeak, Pmean, PEEP
• Breath Rate
• Rate spn
• FiO2%
• Graph Display : P-t, F-t waveforms, P-V and F-V Loops
Alarm
• High airway pressure / Low airway pressure
• High tidal voleme / Low tidal volume
• High frequency / Low frequency
• High Minute Volume / Low Minute Volume
• High FiO2 / Low FiO2 (should option FiO2 function)
• Apnea
• Power Failure
Vaporizer
• Two vaporizer position (Selectatec with interlock)
• Agent type : Sevoflurane, Halothane, Isoflurane and Enflurane
• Filing type : Pour-fill, Easy-fill, Quick-fill or Key-fill
• Compensation : Flow, pressure and temperature compensation